World Journal of Emergency Medicine, 2022, 13(3): 237-238 doi: 10.5847/wjem.j.1920-8642.2022.035

Pictorial Quiz

Puzzling pacing threshold in a patient with cardiac resynchronization therapy device

Song-yun Chu, Jing Zhou,, Yan-sheng Ding

Department of Cardiology, Peking University First Hospital, Beijing 100034, China

Corresponding authors: Jing Zhou, Email:drzhoujing123@126.com

Received: 2021-06-6   Accepted: 2021-12-26  

PDF (2905KB) Metadata Metrics Related articles Export EndNote| Ris| Bibtex  Favorite

Cite this article

Song-yun Chu, Jing Zhou, Yan-sheng Ding. Puzzling pacing threshold in a patient with cardiac resynchronization therapy device. World Journal of Emergency Medicine, 2022, 13(3): 237-238 doi:10.5847/wjem.j.1920-8642.2022.035

CASE

A man in his sixties with dilated cardiomyopathy was admitted to our hospital for the replacement of his cardiac resynchronization therapy (CRT) device. Although his condition had been initially improved after CRT implantation and he had taken an optimized medication regimen and undergone regular device follow-up, echocardiography showed progressive dyspnea and edema with decreased left ventricular ejection fraction (LVEF). During the procedure, his CRT was disconnected and tested using an analyzer module 2290 of the Medtronic 2090 programmer (Medtronic Inc., USA), and a temporary pacemaker (Model 5318; Medtronic Inc., USA) and high pacing thresholds (>5 V/0.5 ms) were noted for both ventricles (Figures 1 A and B). No abnormality in sensing or lead impedance was detected. No significant change in the location or integrity of the leads was found by fluoroscopy. Pre-procedure interrogation showed regular ventricular signals in the intracardiac electrogram (EGM) after the pacing spikes throughout the output decreased to their minimum values (Figure 1C). Apparently, excellent thresholds were also shown by EGM after connecting the new device and repeated testing.

Figure 1.

Figure 1.   Pacing threshold testing during the procedure (A and B) and before the procedure (C). A and B: right ventricular and left ventricular pacing threshold test during the procedure; arrows indicated the loss of capture; C: “satisfactory” biventricular pacing before and after the replacement procedure.


Questions

Which is the correct pacing threshold? What is the most likely reason for the discrepancy of the pacing thresholds measured before and during the operation?

A. >5.0 V/0.5 ms, dislodgement of the leads

B. <1.0 V/0.5 ms, malfunction of the analyzer

C. >5.0 V/0.5 ms, inappropriate testing mode selection

Answers

C. >5.0 V/0.5 ms, inappropriate testing mode selection

DISCUSSION

Lack of response to CRT warrants exploration for causes. Because the left ventricular (LV) lead is placed within the coronary sinus rather than in the myocardium, the rates of dislodgement and micro-dislodgement are higher.[1] However, it is difficult to explain the increase in right ventricular (RV) thresholds and how they return to normal without any repositioning. Analyzer malfunction is extremely rare,[2] and the measures were confirmed by the temporary pacemaker testing.

In fact, “satisfactory” thresholds were suggested by regular ventricular signals after the pacing spikes in EGM, while no change in QRS morphology in electrocardiogram (ECG) was noted when modulation of the atrioventricular (AV) and interventricular (VV) delay. Pseudocapture was doubted after testing with DDD-mode and short AV delay, in which the atrial signals were sensed and ventricular stimuli were delivered before the intrinsic activation, producing apparently captured beats (Figure 2C). VVI-mode testing, as in the procedure, revealed the dissociation of the stimuli and the ventricular activities. A closer look at the multiple EGMs in DDD-mode showed the long latency between the pacing spikes and the activations. It also showed an intact activation sequence, whichever LV or RV paced (Figures 2 A and B), indicating capture failure. A subtle change occurred in the ventricular EGM (Figures 2 D-G) years earlier and led to the progressive clinical exacerbation.

Figure 2.

Figure 2.   Comparison of the EGM and ECG during the procedure and after the initial implantation. A and B: testing of pacing thresholds of right ventricle and left ventricle with DDD-mode and shortened AV delay; C: the pseudocaptured QRS configuration was identical to the intrinsic LBBB pattern before the CRT implantation; D and E: EGM and ECG immediately after CRT implantation; F: EGM of intrinsic beats; G: so-called biventricular pacing capture. EGM: electrogram; ECG: electrocardiogram; AV: atrioventricular; LBBB: left bundle branch block; CRT: cardiac resynchronization therapy.


In viewing the unacceptably high pacing thresholds for RV and LV, we attempted to implant new leads during the procedure. Angiography showed multiple severe stenosis of the left subclavian to supra vena cava. There was also adhesion between the defibrillation coil and the vessel wall of supra vena cava, which prevented the leads from passing through. After discussing options with the patient, we kept the device acting only as a defibrillator with minimal ventricular pacing because the patient was not pacemaker-dependent. Elective total extraction of the transvenous ventricular leads and re-implantation of the LV and RV leads under general anesthesia with a cardiac surgeon stand-by were suggested. However, the patient refused further invasive procedure and decided to accept medical treatment for heart failure.

Funding: None.

Ethical approval: Not needed.

Conflicts of interest: The authors have no commercial associations or sources of support that might pose a conflict of interest.

Contributors: SYC proposed the study and wrote the first draft. All authors contributed to the design and interpretation of the study and to further drafts.

Reference

Nobre Menezes M, de Sousa J, Marques P.

Coronary vein pacing with standard active fixation leads for cardiac resynchronization

Europace. 2015; 17(10):1562.

DOI:10.1093/europace/euv207      PMID:26142454      [Cited within: 1]

Vinther M, Risum N, Philbert BT.

A large inherent delay between the ECG and EGM signals in the pacing system analyzer from Medtronic makes it unsuitable to estimate timing events during CRT implantation

J Electrocardiol. 2020; 58:33-6.

DOI:10.1016/j.jelectrocard.2019.11.001      URL     [Cited within: 1]

/